Electroanalytical determination of carbendazim and fenamiphos in natural waters using a diamond electrode
Data
2012-07-01
Tipo
Artigo
Título da Revista
ISSN da Revista
Título de Volume
Resumo
In this study, a method for electroanalytical determination of carbendazim (CBZ) and fenamiphos (FNP) in natural and spiked water was developed using square wave voltammetly in Na2HPO4 0.1 mol L-1 as supporting electrolyte. the calibration curve for carbendazim detection presented good linearity in the concentration range of 0.50 to 15.0 mu M, with a sensitivity of 0.080 A/mol L-1 and a linearity of 0.998. the oxidation of fenamiphos on BDD electrode shows a dynamic range of concentration of 0.5 to 25.0 mu M and sensibility of 0.14 A/mol L-1. the recovery experiments showed values between 70 and 100% for spiked samples thus indicating the feasibility of the electroanalytical methodology to quantify CBZ and FNP in pure or natural waters. (C) 2012 Elsevier B.V. All rights reserved.
Descrição
Citação
Diamond and Related Materials. Lausanne: Elsevier Science Sa, v. 27-28, p. 54-59, 2012.